Overview

In order to allow SAXS to pass through while measuring WAXS, scientists have been asking for an X-ray detector with a hole in the middle. The resulting LX series has a long, rectangular detecting surface, engineered with a notch to allow the direct beam and SAXS data to pass through. Designed with the new, exclusive Rayonix chip utilizing Frame Transfer Technology.

Technical Specifications

Technical Specifications

Type 3 FT-CCDs bonded to fiber-optic tapers
Sensor Proprietary Rayonix frame-transfer CCD
Fiber-optic Taper Demagnification Ratio 2.92:1
Phosphor standard 40μm thick, custom options available
PSF (FWHM) Depends on Phosphor Option Chosen 100μm with standard phosphor, 65μm with 25μm phosphor
Active Imaging Surface 85mm x 255mm, 21600mm²
Sensor Operating Temperature -80°C
Readout Electronics (# of Channels) 48 channels
Readout Noise High Speed Mode: 8e⁻/pixel Low Noise Mode: 4e⁻/pixel
Dead Time Between Images 1 millisecond
Dark Current 0.003e⁻/pixel/second, or 0.0004 photons/pixel/second (12keV)
DQE 0.8
Sensitivity (Gain) 7e⁻/12keV photon
Full Well Capacity 400ke⁻/pixel (at standard binning, 2 x 2)
Cooling Closed-cycle refrigeration

Technical Specifications
User-configurable Imaging Parameters
Empty section. Edit page to add content here.
Physical Dimensions
Empty section. Edit page to add content here.